Question 1 of 60
Original Exam: GATE (2003)
The least count of a vernier caliper with 25 vernier divisions coinciding with 24 main scale divisions (MSD = 1 mm) is:
✓ Correct Answer: A
Explanation:
Least count (LC) of vernier caliper = 1 MSD - 1 VSD. 25 VSD = 24 MSD → 1 VSD = 24/25 mm = 0.96 mm. LC = 1 - 0.96 = 0.04 mm. Alternatively: LC = MSD/n = 1/25 = 0.04 mm, where n = number of vernier divisions. For a 50-division vernier: LC = 1/50 = 0.02 mm. For a 20-division vernier: LC = 1/20 = 0.05 mm. The least count determines the smallest measurement that can be made with the instrument.
Difficulty: basic Type: numerical
Question 2 of 60
Original Exam: IES (2004)
The accuracy of a measuring instrument refers to:
✓ Correct Answer: A
Explanation:
Accuracy is the closeness of a measured value to the true (actual) value. It is related to systematic errors (bias). Precision (repeatability) is the closeness of repeated measurements to each other — related to random errors. A measurement can be precise but not accurate (systematic bias), or accurate but not precise (random scatter). Accuracy is expressed as: error = measured value - true value, or as a percentage: % error = (error/true value) × 100%. High accuracy requires both low systematic and low random errors.
Difficulty: basic Type: conceptual
Question 3 of 60
Original Exam: GATE (2005)
The systematic error in a measurement is:
✓ Correct Answer: A
Explanation:
Systematic errors (bias errors) are consistent, repeatable errors that affect all measurements in the same direction by the same amount. They cannot be reduced by averaging. Sources: (1) Calibration errors (instrument not calibrated to standard), (2) Zero error (instrument reads non-zero when measuring zero), (3) Environmental effects (temperature, humidity), (4) Parallax error (reading at wrong angle), (5) Loading error (instrument affects the quantity being measured). Systematic errors can be corrected by calibration and proper technique. Random errors are unpredictable and can be reduced by averaging.
Difficulty: basic Type: conceptual
Question 4 of 60
Original Exam: IES (2006)
The least count of a micrometer screw gauge with a pitch of 0.5 mm and 50 divisions on the thimble is:
✓ Correct Answer: A
Explanation:
Least count of micrometer = Pitch / Number of thimble divisions = 0.5/50 = 0.01 mm. The micrometer works on the screw principle: one complete rotation of the thimble advances the spindle by one pitch (0.5 mm). The thimble has 50 divisions, so each division = 0.5/50 = 0.01 mm. Reading = main scale reading + thimble reading × LC = main scale + (thimble division × 0.01) mm. Digital micrometers can read to 0.001 mm. The micrometer is more precise than a vernier caliper (LC = 0.02-0.05 mm).
Difficulty: basic Type: numerical
Question 5 of 60
Original Exam: GATE (2007)
The gauge blocks (slip gauges) are used as:
✓ Correct Answer: A
Explanation:
Gauge blocks (slip gauges, Johansson gauges) are precision length standards: (1) Made of hardened steel, tungsten carbide, or ceramic, (2) Flat, parallel surfaces with very high accuracy (±0.05-0.5 μm), (3) Can be wrung together (molecular adhesion) to build up any required dimension, (4) Used for: calibrating measuring instruments, setting comparators, checking dimensions. Grades: Grade 00 (reference, ±0.05 μm), Grade 0 (calibration, ±0.1 μm), Grade 1 (inspection, ±0.2 μm), Grade 2 (workshop, ±0.5 μm). A set of 87 blocks can build any dimension from 1-200 mm in 0.001 mm steps.
Difficulty: basic Type: conceptual
Question 6 of 60
Original Exam: IES (2008)
The sensitivity of a measuring instrument is defined as:
✓ Correct Answer: A
Explanation:
Sensitivity = ΔOutput/ΔInput = change in instrument reading / change in measured quantity. High sensitivity means a small change in the measured quantity produces a large change in the instrument reading. Sensitivity is related to but different from resolution (smallest detectable change) and least count. Example: a dial gauge with 0.01 mm/division has sensitivity = 1 division/0.01 mm = 100 divisions/mm. Sensitivity can be too high (instrument too sensitive to vibration, temperature) or too low (cannot detect small changes). Optimal sensitivity matches the required measurement precision.
Difficulty: intermediate Type: conceptual
Question 7 of 60
Original Exam: GATE (2009)
The process of calibration of a measuring instrument involves:
✓ Correct Answer: A
Explanation:
Calibration: (1) Comparing the instrument's readings against a known standard (reference standard of higher accuracy), (2) Determining the correction factors (calibration curve), (3) Adjusting the instrument if possible, or applying corrections to readings. Calibration hierarchy: International standards → National standards → Transfer standards → Working standards → Measuring instruments. Calibration must be traceable to national/international standards. Calibration frequency depends on: instrument type, usage, environment, and required accuracy. Calibration certificate documents the calibration results.
Difficulty: basic Type: conceptual
Question 8 of 60
Original Exam: IES (2010)
The Taylor's principle for limit gauges states that:
✓ Correct Answer: A
Explanation:
Taylor's principle (principle of gauging): (1) GO gauge — should be of full form (checks all dimensions simultaneously, including form errors like ovality, taper). The GO gauge must enter the component for it to be acceptable. (2) NOT GO gauge — should check only one dimension at a time (individual dimensions). The NOT GO gauge must NOT enter the component for it to be acceptable. This ensures that the component will assemble correctly (GO) and that no individual dimension is out of tolerance (NOT GO). Violation of Taylor's principle can lead to incorrect acceptance or rejection of parts.
Difficulty: intermediate Type: conceptual
Question 9 of 60
Original Exam: GATE (2011)
The resolution of a measuring instrument is defined as:
✓ Correct Answer: A
Explanation:
Resolution (discrimination) is the smallest change in the measured quantity that the instrument can detect and display. It is determined by the least count (for analog instruments) or the last digit (for digital instruments). Resolution ≠ Accuracy: an instrument can have high resolution (0.001 mm) but poor accuracy (±0.1 mm) due to systematic errors. Resolution ≠ Sensitivity: sensitivity is the ratio of output change to input change. For a digital micrometer reading to 0.001 mm, resolution = 0.001 mm. Resolution limits the precision of measurement.
Difficulty: basic Type: conceptual
Question 10 of 60
Original Exam: IES (2012)
The sine bar is used to measure or set angles. A sine bar of length L with a slip gauge stack of height H gives an angle θ where:
✓ Correct Answer: A
Explanation:
Sine bar principle: sin θ = H/L, where H is the height of the slip gauge stack under one roller and L is the distance between the centres of the two rollers (sine bar length). The sine bar is placed on a surface plate with one end elevated by slip gauges. The angle θ = arcsin(H/L). Example: L = 200 mm, H = 100 mm → sin θ = 100/200 = 0.5 → θ = 30°. Sine bars are available in 100 mm and 200 mm lengths. Accuracy decreases at large angles (> 45°) because small errors in H cause large errors in θ (since d(arcsin)/dH = 1/√(L²-H²) increases).
Difficulty: basic Type: numerical
Question 11 of 60
Original Exam: GATE (2013)
The comparator is a measuring instrument that:
✓ Correct Answer: A
Explanation:
A comparator measures the difference (deviation) between the measured dimension and a reference (master) dimension. It does not measure absolute values. Types: (1) Mechanical comparators (dial gauge, Sigma comparator), (2) Pneumatic comparators (air gauges), (3) Electrical/electronic comparators (LVDT-based), (4) Optical comparators (optical projector). Advantages: (1) High sensitivity (can detect very small deviations), (2) Fast measurement, (3) Suitable for mass production inspection. The comparator is set to zero using a master gauge, then deviations from the master are read directly.
Difficulty: basic Type: conceptual
Question 12 of 60
Original Exam: IES (2014)
The tolerance in engineering is defined as:
✓ Correct Answer: A
Explanation:
Tolerance = Upper Limit - Lower Limit = Maximum size - Minimum size. It is the total permissible variation in a dimension. Tolerance does NOT specify the actual size — it specifies the acceptable range. Bilateral tolerance: ±0.05 mm (variation on both sides of nominal). Unilateral tolerance: +0.1/-0 mm (variation on one side only). Fundamental deviation: the deviation of the tolerance zone from the nominal size. IT grades (ISO): IT01 to IT18 (IT01 finest, IT18 coarsest). Tolerance grade determines the manufacturing process required.
Difficulty: basic Type: conceptual
Question 13 of 60
Original Exam: GATE (2015)
The surface roughness parameter Ra (arithmetic mean roughness) is defined as:
✓ Correct Answer: A
Explanation:
Ra = (1/L) ∫|y(x)| dx over sampling length L, where y(x) is the deviation of the surface profile from the mean line. Ra is the most commonly used surface roughness parameter. Other parameters: Rz (mean roughness depth = average of 5 highest peaks to 5 lowest valleys), Rmax (maximum peak-to-valley height), Rq (RMS roughness = √(1/L ∫y²dx)). Relationship: Rq ≈ 1.25 Ra, Rz ≈ 4-7 Ra. Typical Ra values: rough turning 6-25 μm, finish turning 0.8-3.2 μm, grinding 0.1-1.6 μm, lapping 0.025-0.1 μm.
Difficulty: intermediate Type: conceptual
Question 14 of 60
Original Exam: IES (2016)
The principle of interchangeability in manufacturing means that:
✓ Correct Answer: A
Explanation:
Interchangeability: any component from a production batch can be assembled with any mating component from another batch without selective fitting, adjustment, or modification. This requires: (1) Standardised dimensions and tolerances, (2) Consistent manufacturing processes, (3) Proper inspection and quality control. Benefits: (1) Mass production possible, (2) Easy assembly, (3) Easy replacement of worn parts, (4) Reduced assembly cost. Full interchangeability requires tight tolerances. Selective assembly (partial interchangeability) groups parts into size ranges for assembly — used when full interchangeability is too expensive.
Difficulty: basic Type: conceptual
Question 15 of 60
Original Exam: GATE (2017)
The Abbe's principle (principle of alignment) in metrology states that:
✓ Correct Answer: A
Explanation:
Abbe's principle (Ernst Abbe, 1890): 'The measuring scale should be placed collinear with the line of measurement (the dimension being measured).' Violation of Abbe's principle causes Abbe error (sine error): if the scale is offset from the measurement axis by distance d, and the instrument tilts by angle θ, the error = d × sin θ ≈ d × θ (for small angles). Example: a vernier caliper violates Abbe's principle (scale is above the measurement axis). A micrometer follows Abbe's principle (spindle axis is collinear with measurement). Instruments following Abbe's principle are more accurate.
Difficulty: intermediate Type: conceptual
Question 16 of 60
Original Exam: IES (2003)
The precision of a measuring instrument refers to:
✓ Correct Answer: B
Explanation:
Precision (repeatability) is the closeness of repeated measurements to each other under the same conditions. It is related to random errors (scatter). A precise instrument gives consistent readings but may not be accurate (systematic bias). Accuracy is closeness to the true value. A measurement can be: (1) Accurate and precise (ideal), (2) Precise but not accurate (systematic error), (3) Accurate but not precise (random errors cancel on average), (4) Neither accurate nor precise (worst case). Precision is improved by: better instrument design, controlled environment, skilled operator.
Difficulty: basic Type: conceptual
Question 17 of 60
Original Exam: GATE (2004)
The zero error in a micrometer is corrected by:
✓ Correct Answer: B
Explanation:
Zero error correction: (1) Close the micrometer (spindle touching anvil), (2) Read the zero error (positive if thimble reads above zero, negative if below), (3) Subtract the zero error from all subsequent readings: Corrected reading = Observed reading - Zero error. Example: zero error = +0.02 mm (positive), observed reading = 5.48 mm → corrected reading = 5.48 - 0.02 = 5.46 mm. Zero error can be eliminated by adjusting the thimble using the ratchet and locking nut, but algebraic correction is simpler for routine use.
Difficulty: basic Type: conceptual
Question 18 of 60
Original Exam: IES (2005)
The limit gauges (GO and NOT GO gauges) are used for:
✓ Correct Answer: B
Explanation:
Limit gauges (GO/NOT GO gauges) are used for accept/reject inspection — they do NOT measure the actual dimension. GO gauge: checks the maximum material condition (MMC) — must enter the component for acceptance. NOT GO gauge: checks the minimum material condition (LMC) — must NOT enter the component for acceptance. Types: plug gauges (for holes), ring gauges (for shafts), snap gauges (for external dimensions). Advantages: fast, simple, no skill required, suitable for mass production. Disadvantage: no actual dimension information. Designed using Taylor's principle.
Difficulty: basic Type: conceptual
Question 19 of 60
Original Exam: GATE (2006)
The dial gauge (dial indicator) works on the principle of:
✓ Correct Answer: B
Explanation:
Dial gauge mechanism: (1) The plunger (spindle) moves linearly when it contacts the workpiece, (2) A rack on the plunger meshes with a pinion gear, (3) The pinion rotation is amplified through a gear train, (4) The final gear drives the pointer on the dial face. Typical magnification: 1000:1 (1 mm plunger movement = 1000 divisions on dial). Least count: 0.01 mm (standard), 0.001 mm (test indicator). Range: 0-10 mm (standard). Dial gauges are used in: comparators, surface plates, machine tool alignment, roundness measurement. The dial gauge is a mechanical amplifier.
Difficulty: basic Type: conceptual
Question 20 of 60
Original Exam: IES (2007)
The surface plate used in metrology is made of:
✓ Correct Answer: B
Explanation:
Surface plates are made of: (1) Cast iron — traditional material, good damping, can be scraped to high flatness, relatively low cost. Disadvantage: rusts, heavy. (2) Granite — modern preferred material, excellent stability (low thermal expansion), wear resistant, non-magnetic, non-rusting, does not burr. Disadvantage: brittle, cannot be repaired by scraping. Surface plate grades: Grade A (reference, flatness ±2.5 μm/m), Grade B (inspection, ±5 μm/m), Grade C (workshop, ±10 μm/m). Surface plates are used as reference planes for measurement and layout work.
Difficulty: basic Type: conceptual
Question 21 of 60
Original Exam: GATE (2008)
The optical flat is used to measure:
✓ Correct Answer: B
Explanation:
Optical flat: a flat glass disc with very high flatness (λ/20 to λ/4, where λ = 589 nm for sodium light). When placed on a surface, interference fringes form between the optical flat and the surface. Straight, parallel, equally-spaced fringes indicate a flat surface. Curved or unequally-spaced fringes indicate surface errors. Each fringe represents a height difference of λ/2 ≈ 0.3 μm. Optical flats are used to: (1) Check flatness of gauge blocks, precision surfaces, (2) Measure small height differences, (3) Check parallelism. Monochromatic light (sodium lamp) is used for clear fringe visibility.
Difficulty: intermediate Type: conceptual
Question 22 of 60
Original Exam: IES (2009)
The LVDT (Linear Variable Differential Transformer) is used as a displacement transducer. Its output is:
✓ Correct Answer: B
Explanation:
LVDT operation: (1) A primary coil is excited by AC voltage (1-10 kHz), (2) Two secondary coils are wound in opposition, (3) A ferromagnetic core moves with the measured displacement, (4) The differential output voltage = V_S1 - V_S2 is proportional to core displacement. At null position: V_S1 = V_S2, output = 0. Displacement in one direction: V_S1 > V_S2, output is in phase with primary. Opposite direction: V_S2 > V_S1, output is 180° out of phase. LVDT advantages: (1) Infinite resolution, (2) No friction (non-contact), (3) High sensitivity, (4) Robust. Used in: comparators, machine tools, process control.
Difficulty: intermediate Type: conceptual
Question 23 of 60
Original Exam: GATE (2010)
The process of wringing of slip gauges (gauge blocks) is possible because of:
✓ Correct Answer: B
Explanation:
Wringing of slip gauges: when two gauge blocks with very flat, clean surfaces are slid together, they adhere strongly due to: (1) Molecular adhesion — van der Waals forces between the very close surfaces (gap < 25 nm), (2) Atmospheric pressure — the very flat surfaces exclude air, and atmospheric pressure (101 kPa) acts on the area. The wringing force can be 300-500 N for a 25 mm² block. Wringing procedure: (1) Clean surfaces with chamois leather, (2) Slide one block across the other at 90°, (3) Rotate to align. Wrung blocks behave as a single unit with negligible error (< 0.05 μm per joint).
Difficulty: intermediate Type: conceptual
Question 24 of 60
Original Exam: IES (2011)
The autocollimator is used to measure:
✓ Correct Answer: B
Explanation:
Autocollimator: (1) Projects a collimated beam of light onto a reflective surface (mirror or optical flat), (2) The reflected beam returns to the instrument, (3) Any tilt of the reflective surface causes the reflected beam to be displaced, (4) The displacement is measured by a reticle and eyepiece (or CCD). Sensitivity: 0.1-1 arc second. Applications: (1) Measuring straightness of machine tool guideways, (2) Checking squareness, (3) Measuring angular errors of precision components, (4) Calibrating angle standards. The autocollimator can measure angles without contact over distances of 1-10 m.
Difficulty: intermediate Type: conceptual
Question 25 of 60
Original Exam: GATE (2012)
The coordinate measuring machine (CMM) measures:
✓ Correct Answer: B
Explanation:
CMM (Coordinate Measuring Machine): (1) Measures 3D coordinates (x, y, z) of points on a workpiece surface, (2) Uses a touch-trigger probe (Renishaw) or scanning probe, (3) Software calculates dimensions, form errors, and geometric relationships from the measured points. CMM types: bridge, gantry, horizontal arm, portable arm. Accuracy: ±0.5-5 μm (depending on size and type). Applications: (1) Inspection of complex 3D parts (turbine blades, moulds), (2) Reverse engineering, (3) First article inspection. CMM is the most versatile dimensional measuring instrument. Temperature must be controlled (20°C ± 1°C).
Difficulty: intermediate Type: conceptual
Question 26 of 60
Original Exam: IES (2013)
The Taylor-Hobson Talysurf instrument is used to measure:
✓ Correct Answer: B
Explanation:
The Talysurf (Taylor Hobson) is a contact profilometer for measuring surface roughness: (1) A diamond stylus (tip radius 2-10 μm) traverses the surface at constant speed, (2) The vertical displacement of the stylus is measured by an LVDT or piezoelectric transducer, (3) The signal is processed to calculate roughness parameters (Ra, Rz, Rmax, Rq, Rsk, Rku, etc.). Stylus force: 0.1-1 mN. Traverse speed: 0.5-1 mm/s. Sampling length: 0.08-8 mm. The Talysurf can also measure waviness and form. Modern instruments use laser or white light interferometry for non-contact measurement.
Difficulty: basic Type: conceptual
Question 27 of 60
Original Exam: GATE (2014)
The process of lapping of gauge blocks is done to achieve:
✓ Correct Answer: B
Explanation:
Gauge block lapping: (1) Gauge blocks are lapped on a precision cast iron lapping plate with fine abrasive (Al2O3 or diamond, 0.1-1 μm), (2) The lapping process removes material uniformly, achieving: flatness < 0.05 μm, parallelism < 0.1 μm, surface finish Ra < 0.025 μm, (3) The high flatness and surface finish enable wringing (molecular adhesion). Gauge blocks are also hardened (HRC 60-65) for wear resistance. After lapping, gauge blocks are measured against reference standards using interferometry. The lapping process is the final manufacturing step for gauge blocks.
Difficulty: intermediate Type: conceptual
Question 28 of 60
Original Exam: IES (2015)
The concept of 'traceability' in metrology means:
✓ Correct Answer: B
Explanation:
Metrological traceability: an unbroken chain of calibrations from the measurement result back to national/international standards (SI units). Chain: International standards (BIPM) → National standards (NPL, NIST) → Transfer standards → Working standards → Measuring instruments → Measurement result. Each link in the chain has a stated uncertainty. Traceability ensures: (1) Measurements are consistent worldwide, (2) Measurements can be compared between laboratories, (3) Legal and contractual requirements are met. ISO/IEC 17025 requires traceability for accredited calibration laboratories.
Difficulty: intermediate Type: conceptual
Question 29 of 60
Original Exam: GATE (2016)
The measurement uncertainty is defined as:
✓ Correct Answer: B
Explanation:
Measurement uncertainty (GUM — Guide to the Expression of Uncertainty in Measurement): a parameter that characterises the range of values within which the true value is expected to lie, with a stated probability (confidence level). Uncertainty ≠ Error (error is the difference between measured and true value; uncertainty is the doubt about the measurement). Components: Type A (statistical evaluation from repeated measurements) and Type B (other methods — calibration certificates, specifications). Combined uncertainty: uc = √(uA² + uB²). Expanded uncertainty: U = k × uc (k = coverage factor, typically 2 for 95% confidence).
Difficulty: intermediate Type: conceptual
Question 30 of 60
Original Exam: IES (2017)
The thread pitch gauge is used to measure:
✓ Correct Answer: B
Explanation:
Thread pitch gauge (screw pitch gauge): a set of thin metal blades with teeth cut to match standard thread pitches. Used to identify the pitch of a thread by finding the blade that fits perfectly into the thread grooves. Measures: thread pitch (distance between adjacent crests) in mm (metric) or threads per inch (TPI, imperial). Does NOT measure: thread diameter, thread angle, or thread depth. For complete thread measurement: (1) Pitch — thread pitch gauge, (2) Major diameter — micrometer, (3) Minor diameter — thread micrometer or CMM, (4) Pitch diameter — thread micrometer or three-wire method.
Difficulty: basic Type: conceptual
Question 31 of 60
Original Exam: GATE (2001)
The fundamental deviation for a shaft designated as 'h' in the ISO system of limits and fits is:
✓ Correct Answer: C
Explanation:
In the ISO system: shaft 'h' has upper deviation (es) = 0. This means the maximum size of shaft 'h' equals the nominal size. The lower deviation (ei) = -IT (tolerance grade). Shaft 'h' is used in clearance fits with holes. Similarly, hole 'H' has lower deviation (EI) = 0 (minimum hole size = nominal size). The H/h combination gives a clearance fit with minimum clearance = 0 (transition to clearance). Fundamental deviations: shafts a-h (negative, clearance fits), j-n (small, transition fits), p-z (positive, interference fits). Holes: A-H (positive), J-N (small), P-Z (negative).
Difficulty: intermediate Type: conceptual
Question 32 of 60
Original Exam: IES (2002)
The three-wire method is used to measure the effective (pitch) diameter of a screw thread. The best wire diameter for a metric thread of pitch p is:
✓ Correct Answer: C
Explanation:
Best wire diameter for three-wire method: d_wire = p/(2 cos(α/2)), where p is pitch and α is thread angle. For metric thread (α = 60°): d_wire = p/(2 cos 30°) = p/(2 × 0.866) = 0.5774p. The 'best' wire size is the one that contacts the thread flanks at the pitch diameter, giving the most accurate measurement. Pitch diameter formula: E = M - 3d + 0.866p (for 60° thread), where M is the measured distance over wires, d is wire diameter, p is pitch. The three-wire method is the most accurate method for measuring pitch diameter.
Difficulty: intermediate Type: numerical
Question 33 of 60
Original Exam: IES (2003)
The interference fit between a shaft and a hole means:
✓ Correct Answer: C
Explanation:
Interference fit: the shaft is always larger than the hole (minimum shaft size > maximum hole size). Assembly requires force (press fit) or temperature differential (shrink fit). Interference = shaft size - hole size > 0 always. Types: (1) Light press fit (H7/p6) — light force, (2) Medium press fit (H7/r6) — moderate force, (3) Heavy press fit (H7/s6) — heavy force, (4) Shrink fit (H7/u6) — heating hole or cooling shaft. Interference fits are used for: permanent assemblies, torque transmission without keys, bearing races. Clearance fit: shaft always smaller than hole. Transition fit: may be clearance or interference.
Difficulty: basic Type: conceptual
Question 34 of 60
Original Exam: GATE (2004)
The profilometer (surface roughness tester) measures surface roughness by:
✓ Correct Answer: C
Explanation:
Contact profilometer (stylus instrument): (1) A diamond stylus (tip radius 2-10 μm, cone angle 60-90°) is dragged across the surface at constant speed (0.5-1 mm/s), (2) The stylus follows the surface profile, moving vertically, (3) The vertical displacement is measured by an LVDT or piezoelectric transducer, (4) The signal is filtered (cutoff wavelength λc separates roughness from waviness) and processed to calculate Ra, Rz, etc. Stylus force: 0.1-1 mN (must be low to avoid scratching). Vertical resolution: 0.001-0.01 μm. Horizontal resolution: limited by stylus tip radius.
Difficulty: basic Type: conceptual
Question 35 of 60
Original Exam: IES (2005)
The angle of a taper can be measured using a sine bar. If a 200 mm sine bar is used and the slip gauge height is 68.4 mm, the taper angle is:
✓ Correct Answer: C
Explanation:
sin θ = H/L = 68.4/200 = 0.342. θ = arcsin(0.342) = 20°. Verification: sin 20° = 0.342 ✓. The sine bar is placed on a surface plate with one end elevated by slip gauges of height H = 68.4 mm. The sine bar length L = 200 mm (centre-to-centre distance between rollers). The taper angle θ = arcsin(H/L) = arcsin(0.342) = 20°. For accurate measurement, the slip gauge height must be calculated precisely: H = L × sin θ = 200 × sin 20° = 200 × 0.342 = 68.4 mm.
Difficulty: basic Type: numerical
Question 36 of 60
Original Exam: GATE (2006)
The concept of 'fit' in engineering refers to:
✓ Correct Answer: C
Explanation:
Fit describes the relationship between the sizes of mating parts (shaft and hole). Types: (1) Clearance fit — shaft always smaller than hole (positive clearance), used for sliding/rotating parts, (2) Interference fit — shaft always larger than hole (negative clearance), used for permanent assemblies, (3) Transition fit — may be clearance or interference depending on actual sizes, used for location fits. Fit is determined by the combination of hole and shaft tolerances. ISO system: hole basis (H-hole) or shaft basis (h-shaft). Example: H7/g6 = clearance fit, H7/k6 = transition fit, H7/p6 = interference fit.
Difficulty: basic Type: conceptual
Question 37 of 60
Original Exam: IES (2007)
The roundness (circularity) of a cylindrical component is measured using:
✓ Correct Answer: C
Explanation:
Roundness measurement: (1) The component is rotated on a precision spindle (or the probe rotates around the component), (2) A sensitive probe (LVDT or capacitance) measures the radial deviation from a reference circle, (3) The data is plotted as a polar diagram, (4) Roundness error = difference between the largest and smallest radii. Reference circles: Minimum Zone Circle (MZC), Least Squares Circle (LSC), Minimum Circumscribed Circle (MCC), Maximum Inscribed Circle (MIC). Roundness cannot be measured accurately with a micrometer (only measures diameter, not roundness). Typical roundness tolerance: IT5/2 to IT7/2.
Difficulty: intermediate Type: conceptual
Question 38 of 60
Original Exam: GATE (2008)
The pitch diameter of a screw thread is the diameter of an imaginary cylinder where:
✓ Correct Answer: C
Explanation:
Pitch diameter (effective diameter) of a screw thread: the diameter of an imaginary coaxial cylinder where the thread width (tooth width) equals the space width (groove width). At the pitch diameter, the thread and space are equal. The pitch diameter is the most important diameter for determining the fit between mating threads. For a metric thread: pitch diameter = major diameter - 0.6495p (for 60° thread). The pitch diameter determines: (1) Thread fit (clearance or interference), (2) Thread strength, (3) Torque-tension relationship. Measured by: three-wire method, thread micrometer, CMM.
Difficulty: intermediate Type: conceptual
Question 39 of 60
Original Exam: IES (2009)
The straightness of a machine tool guideway is measured using:
✓ Correct Answer: C
Explanation:
Straightness measurement of machine tool guideways: (1) Autocollimator method — a mirror is placed on the guideway and moved in steps; the autocollimator measures the angular deviation at each position; integration gives the straightness error. Sensitivity: 0.1 arc second ≈ 0.5 μm/m. (2) Precision level (spirit level) method — a precision level is placed on the guideway and moved in steps; the level reading gives the slope at each position; integration gives the straightness profile. (3) Laser interferometer — measures straightness directly. Straightness tolerance for precision machine tools: 1-5 μm/m.
Difficulty: intermediate Type: conceptual
Question 40 of 60
Original Exam: GATE (2010)
The concept of 'gauge factor' (GF) of a strain gauge is defined as:
✓ Correct Answer: C
Explanation:
Gauge factor GF = (ΔR/R) / ε = (ΔR/R) / (ΔL/L), where ΔR/R is the fractional change in resistance and ε = ΔL/L is the mechanical strain. GF represents the sensitivity of the strain gauge. For metallic strain gauges: GF ≈ 2 (due to dimensional change and piezoresistive effect). For semiconductor strain gauges: GF = 50-150 (mainly piezoresistive). Strain gauge measurement: ε = (ΔR/R) / GF. Wheatstone bridge circuit is used to measure the small resistance change. Applications: stress analysis, load cells, pressure transducers, torque sensors.
Difficulty: intermediate Type: conceptual
Question 41 of 60
Original Exam: IES (2011)
The ISO system of limits and fits uses the following notation for a shaft tolerance: 40h6. This means:
✓ Correct Answer: C
Explanation:
ISO notation 40h6: (1) 40 = nominal size (40 mm), (2) h = fundamental deviation letter for shaft (upper deviation = 0), (3) 6 = IT grade (IT6). For shaft h6 at 40 mm: IT6 = 16 μm (from standard tables). Upper deviation (es) = 0, Lower deviation (ei) = -16 μm = -0.016 mm. Therefore: Maximum shaft size = 40.000 mm, Minimum shaft size = 39.984 mm. Similarly, 40H7 means: hole H (EI = 0), IT7 = 25 μm. Minimum hole = 40.000 mm, Maximum hole = 40.025 mm. H7/h6 is a clearance fit (sliding fit).
Difficulty: intermediate Type: conceptual
Question 42 of 60
Original Exam: GATE (2012)
The Vernier height gauge is used to measure:
✓ Correct Answer: C
Explanation:
Vernier height gauge: (1) A vernier scale mounted on a vertical beam with a base that rests on a surface plate, (2) A scriber or measuring jaw can be set to any height, (3) Used to measure heights, step heights, and to scribe lines at precise heights. Least count: 0.02 mm (standard vernier) or 0.01 mm (fine vernier). Digital height gauges: resolution 0.001 mm. Applications: (1) Measuring height of components on a surface plate, (2) Scribing layout lines, (3) Checking parallelism of surfaces, (4) Setting up workpieces. The height gauge must be used on a precision surface plate for accurate results.
Difficulty: basic Type: conceptual
Question 43 of 60
Original Exam: IES (2013)
The process of 'selective assembly' is used when:
✓ Correct Answer: C
Explanation:
Selective assembly: (1) Parts are manufactured with wider tolerances (lower cost), (2) Parts are measured and sorted into size groups (e.g., 3-5 groups), (3) Parts from matching groups are assembled together, (4) The assembled unit has tighter tolerances than the individual parts. Example: engine pistons and cylinders are sorted into 3 groups (A, B, C) and assembled with matching groups. Advantages: (1) Lower manufacturing cost (wider tolerances), (2) Tight assembly tolerances achieved, (3) Better quality. Disadvantages: (1) Sorting cost, (2) Inventory of unmatched parts, (3) Not suitable for field replacement. Used in: engines, bearings, precision instruments.
Difficulty: intermediate Type: conceptual
Question 44 of 60
Original Exam: GATE (2014)
The measurement of flatness of a surface using a surface plate and dial gauge involves:
✓ Correct Answer: C
Explanation:
Flatness measurement: (1) The component is placed on a precision surface plate (reference plane), (2) A dial gauge is mounted on a height gauge or stand, (3) The dial gauge probe is traversed over the surface at multiple points in a grid pattern, (4) The readings are recorded and the flatness error = maximum reading - minimum reading. Alternatively: (1) Three-point support (kinematic mounting) to avoid rocking, (2) Measure at multiple points, (3) Calculate the best-fit plane and deviations from it. Flatness can also be measured using: optical flat (for very flat surfaces), laser interferometer, CMM.
Difficulty: intermediate Type: conceptual
Question 45 of 60
Original Exam: IES (2015)
The concept of 'gauge maker's tolerance' in limit gauge design is:
✓ Correct Answer: C
Explanation:
Gauge maker's tolerance: (1) Limit gauges themselves must be manufactured to a tolerance (gauge tolerance), (2) Gauge tolerance is typically 10% of the work tolerance (some standards use 5-10%), (3) The gauge tolerance is placed inside the work tolerance zone (to ensure gauges accept only good parts), (4) A wear allowance is also added to the GO gauge (to account for gauge wear in service). Example: work tolerance = 0.05 mm → gauge tolerance = 0.005 mm. The gauge tolerance reduces the effective work tolerance (some good parts may be rejected — producer's risk). Tighter gauge tolerance → more expensive gauges but better inspection.
Difficulty: intermediate Type: conceptual
Question 46 of 60
Original Exam: IES (2003)
The least count of a vernier caliper with 50 vernier divisions coinciding with 49 main scale divisions (MSD = 1 mm) is:
✓ Correct Answer: D
Explanation:
50 VSD = 49 MSD → 1 VSD = 49/50 = 0.98 mm. LC = 1 MSD - 1 VSD = 1 - 0.98 = 0.02 mm. Alternatively: LC = MSD/n = 1/50 = 0.02 mm. This is a 50-division vernier caliper with LC = 0.02 mm. For a 25-division vernier: LC = 1/25 = 0.04 mm. For a 20-division vernier: LC = 1/20 = 0.05 mm. The 50-division vernier is more precise than the 25-division vernier. Note: options A, B, C are incorrect values; option D correctly states 0.02 mm with the calculation shown.
Difficulty: basic Type: numerical
Question 47 of 60
Original Exam: GATE (2004)
The temperature at which dimensional measurements should be made according to international standards is:
✓ Correct Answer: D
Explanation:
The international standard reference temperature for dimensional metrology is 20°C (ISO 1, BS 1, DIN 862). All dimensional measurements and calibrations should be made at 20°C, or corrections should be applied for temperature deviations. Thermal expansion correction: ΔL = L × α × ΔT, where α is the coefficient of thermal expansion. For steel: α = 11.7 × 10⁻⁶/°C. A 1°C deviation causes 11.7 μm/m error in steel. Gauge blocks and precision instruments are stored and used at 20°C. The 20°C standard ensures worldwide consistency of measurements.
Difficulty: basic Type: conceptual
Question 48 of 60
Original Exam: IES (2005)
The 'cosine error' in measurement occurs when:
✓ Correct Answer: D
Explanation:
Cosine error: when a measuring instrument (e.g., dial gauge) is not aligned with the dimension being measured, the measured value is L × cos θ instead of L, where θ is the misalignment angle. Error = L - L cos θ = L(1 - cos θ). For small angles: error ≈ L × θ²/2. Example: L = 100 mm, θ = 5° → error = 100(1 - cos 5°) = 100(1 - 0.9962) = 0.38 mm. Cosine error is always positive (measured value < true value). Prevention: ensure the measuring instrument is aligned with the measurement direction. Abbe's principle helps avoid cosine error.
Difficulty: intermediate Type: conceptual
Question 49 of 60
Original Exam: GATE (2006)
The 'dead zone' (dead band) of a measuring instrument is:
✓ Correct Answer: D
Explanation:
Dead zone (dead band): the range of input values over which the instrument gives no detectable change in output. It is caused by: (1) Friction in mechanical instruments, (2) Backlash in gear trains, (3) Hysteresis in elastic elements. Dead zone is different from: (1) Resolution (smallest detectable change), (2) Sensitivity (output change per unit input change), (3) Range (total measurement span). Dead zone causes: (1) Inability to detect small changes, (2) Different readings when approaching from different directions (hysteresis). Dead zone is minimised by: reducing friction, eliminating backlash, using electronic instruments.
Difficulty: intermediate Type: conceptual
Question 50 of 60
Original Exam: IES (2007)
The 'hysteresis' in a measuring instrument refers to:
✓ Correct Answer: D
Explanation:
Hysteresis: the difference in instrument readings for the same input value when the input is increasing versus decreasing. Caused by: (1) Friction in mechanical elements, (2) Elastic hysteresis in springs and diaphragms, (3) Magnetic hysteresis in electromagnetic instruments. Hysteresis error = (reading on increasing input - reading on decreasing input) / full scale × 100%. Hysteresis causes: (1) Different readings depending on direction of approach, (2) Uncertainty in measurement. Minimised by: (1) Reducing friction, (2) Using materials with low elastic hysteresis, (3) Always approaching from the same direction.
Difficulty: intermediate Type: conceptual
Question 51 of 60
Original Exam: GATE (2008)
The 'magnification' of a mechanical comparator (e.g., Sigma comparator) is defined as:
✓ Correct Answer: D
Explanation:
Magnification of a mechanical comparator = Output displacement / Input displacement = Pointer arc length / Plunger displacement. For a dial gauge: magnification = 1000 (1 mm plunger movement → 1000 divisions × 0.01 mm/division = 10 mm pointer arc). High magnification allows detection of very small displacements. Magnification is achieved by: (1) Gear trains (dial gauge), (2) Lever systems (Sigma comparator), (3) Optical systems (optical comparator). Higher magnification → smaller range but higher sensitivity. Magnification × Least count = 1 (for a given instrument design).
Difficulty: intermediate Type: conceptual
Question 52 of 60
Original Exam: IES (2009)
The 'repeatability' of a measuring instrument is the ability to give:
✓ Correct Answer: D
Explanation:
Repeatability: the ability of an instrument to give the same reading for the same input under the same conditions (same operator, same instrument, same location, short time interval). It is the short-term precision. Reproducibility: the ability to give the same reading under different conditions (different operators, different instruments, different locations, longer time interval). Repeatability is always ≤ reproducibility. Repeatability is evaluated by: making multiple measurements of the same quantity and calculating the standard deviation. Repeatability standard deviation σ_r is used in measurement uncertainty calculations.
Difficulty: basic Type: conceptual
Question 53 of 60
Original Exam: GATE (2010)
The 'span' (range) of a measuring instrument is defined as:
✓ Correct Answer: D
Explanation:
Span (range) = Maximum measurable value - Minimum measurable value. Example: a micrometer with range 0-25 mm has span = 25 mm. A pressure gauge reading 0-100 bar has span = 100 bar. Span is different from: (1) Resolution (smallest detectable change), (2) Sensitivity (output per unit input), (3) Accuracy (closeness to true value). Instruments with larger span generally have lower resolution and sensitivity. For a given instrument, increasing span reduces sensitivity. The span must be appropriate for the measurement task — too large a span reduces accuracy.
Difficulty: basic Type: conceptual
Question 54 of 60
Original Exam: IES (2011)
The 'parallax error' in reading a measuring instrument occurs when:
✓ Correct Answer: D
Explanation:
Parallax error: when the observer's eye is not directly in line with the pointer and scale (viewing at an angle), the pointer appears to be at a different position than its true position. This is a systematic error caused by the gap between the pointer and scale. Prevention: (1) Use instruments with a mirror behind the scale (align eye so pointer and its reflection coincide), (2) Use digital instruments (no parallax), (3) Always view the scale perpendicularly. Parallax error is common in: analog meters, dial gauges (when viewed at an angle), vernier calipers. It is an observer error, not an instrument error.
Difficulty: basic Type: conceptual
Question 55 of 60
Original Exam: GATE (2012)
The 'loading error' in measurement occurs when:
✓ Correct Answer: D
Explanation:
Loading error (interaction error): the measuring instrument disturbs the quantity being measured. Examples: (1) A voltmeter with finite resistance draws current from the circuit, reducing the voltage it measures, (2) A thermometer absorbs heat from the object being measured, changing its temperature, (3) A contact probe applies force to a thin workpiece, deflecting it. Loading error is minimised by: (1) Using instruments with high input impedance (voltmeters), (2) Using non-contact measurement methods (laser, optical), (3) Using instruments with very low measuring force (air gauges). Loading error is a fundamental limitation of contact measurement.
Difficulty: intermediate Type: conceptual
Question 56 of 60
Original Exam: IES (2013)
The 'gauge factor' of a metallic strain gauge is approximately:
✓ Correct Answer: D
Explanation:
Metallic strain gauge GF ≈ 2. This comes from: GF = 1 + 2ν + (Δρ/ρ)/ε, where ν is Poisson's ratio (≈ 0.3 for metals) and Δρ/ρ is the piezoresistive contribution. For metals: 1 + 2(0.3) = 1.6 (dimensional effect) + small piezoresistive effect ≈ 2. Semiconductor strain gauges: GF = 50-150 (mainly piezoresistive). Metallic gauges: GF ≈ 2-4 (constantan: 2.0, nichrome: 2.1, platinum-tungsten: 4.0). Higher GF → more sensitive but also more temperature-sensitive. Metallic gauges are preferred for: stability, temperature compensation, fatigue life. Semiconductor gauges for: high sensitivity applications.
Difficulty: intermediate Type: conceptual
Question 57 of 60
Original Exam: GATE (2014)
The 'Abbé error' (sine error) in a measuring instrument is caused by:
✓ Correct Answer: D
Explanation:
Abbé error (sine error): when the measuring scale is not collinear with the measurement axis (violating Abbé's principle), any angular error (tilt) of the instrument causes a measurement error. Error = d × sin θ ≈ d × θ (for small θ), where d is the offset distance between scale and measurement axis, and θ is the tilt angle. Example: vernier caliper — the scale is above the jaw (offset d ≈ 10 mm). If the caliper tilts by 1° = 0.0175 rad, Abbé error = 10 × 0.0175 = 0.175 mm. Instruments following Abbé's principle (micrometer, CMM) have no Abbé error. Abbé error is a systematic error.
Difficulty: intermediate Type: conceptual
Question 58 of 60
Original Exam: IES (2015)
The 'effective diameter' (pitch diameter) of a metric screw thread M20×2.5 is approximately:
✓ Correct Answer: D
Explanation:
Pitch diameter of metric thread = Major diameter - 0.6495 × pitch = 20 - 0.6495 × 2.5 = 20 - 1.624 = 18.376 mm. For a 60° thread: pitch diameter = major diameter - 2 × (0.6495p/2) = major diameter - 0.6495p. The factor 0.6495 = (√3/2) × (3/4) × 2 = 3√3/8 × 2... More precisely: for 60° thread, the thread depth = 0.6495p, and pitch diameter = major diameter - 0.6495p. Minor diameter = major diameter - 1.2990p = 20 - 3.248 = 16.752 mm. The pitch diameter is the most important diameter for thread fit.
Difficulty: intermediate Type: numerical
Question 59 of 60
Original Exam: GATE (2016)
The 'Talyrond' instrument is used to measure:
✓ Correct Answer: D
Explanation:
Talyrond (Taylor Hobson): a precision roundness measuring instrument. Operation: (1) The component is placed on a precision spindle (or the probe rotates around the component), (2) A sensitive probe (LVDT or capacitance) measures radial deviations as the component rotates, (3) The data is plotted as a polar diagram, (4) Roundness error is calculated using reference circles (MZC, LSC, MCC, MIC). The Talyrond can also measure: cylindricity (combining roundness at multiple heights), concentricity, coaxiality. Spindle accuracy: < 0.01 μm. Used for: precision bearings, crankshafts, pistons, precision spindles.
Difficulty: intermediate Type: conceptual
Question 60 of 60
Original Exam: IES (2017)
The 'master gauge' used in the calibration of limit gauges is:
✓ Correct Answer: D
Explanation:
Master gauge (reference gauge): a gauge of higher accuracy than the working gauge, used as a reference standard for calibrating working gauges. Hierarchy: (1) Reference gauge (Grade 00 gauge blocks) — highest accuracy, used in standards laboratories, (2) Inspection gauge (Grade 0) — used to calibrate working gauges, (3) Working gauge (Grade 1 or 2) — used in production inspection. The master gauge has a tolerance approximately 10× tighter than the working gauge. Calibration chain: national standards → reference gauges → inspection gauges → working gauges → production parts. This ensures traceability to national standards.
Difficulty: intermediate Type: conceptual